4.1
The ADVIA Centaur BRAHMS PCT assay (Siemens Healthcare Diagnostics) is an automated chemiluminescent assay for the determination of procalcitonin in human serum and plasma. The assay principle combines a sandwich immunoassay with the detection of light emission as the final step. It is for use with the ADVIA Centaur/XP and ADVIA Centaur CP analysers. It has a measuring range of 0.02 to 75 nanograms per millilitre, a functional sensitivity of less than 0.05 nanograms per millilitre and an analytical sensitivity of less than 0.02 nanograms per millilitre. The time to result is 26 to 29 minutes, depending on the selected analyser.