PLGF-based testing to help diagnose suspected preterm pre-eclampsia (update of DG23)
Closed for comments This consultation ended on at Request commenting lead permission
3 Committee discussion
The diagnostics advisory committee considered evidence from several sources on the BRAHMS sFlt-1 Kryptor/BRAHMS PLGF plus Kryptor PE ratio, DELFIA Xpress PLGF 1‑2‑3 test, DELFIA Xpress sFlt-1/PLGF ratio, Elecsys immunoassay sFlt‑1/PLGF ratio and Triage PLGF Test. Evidence was considered from the diagnostics assessment report and an overview of that report, and the decision support unit's (DSU) report and updated model. Full details of all the evidence are in the project documents for this guidance.
PLGF-based tests are likely to substantially benefit women with suspected preterm pre-eclampsia
3.1 A patient expert explained that pregnancy can be a particularly worrying time for expectant mothers if they had pre-eclampsia or complications from hypertension in a previous pregnancy. Placental growth factor (PLGF)-based testing can reassure pregnant women with hypertension anxious about complications and risks to the baby and themselves, and increase their confidence in treatment plans. A patient expert highlighted an Action on Pre-eclampsia report that stated that women from African, Asian or Caribbean family backgrounds have a higher risk of developing pre-eclampsia and that PLGF tests may particularly benefit higher risk groups. Clinical experts said that the tests can improve risk assessment and enable early planning for a safe birth. They said they may also help avoid stressful last minute medical interventions. Early planning for at-risk pregnancies also means women at centres without facilities for preterm baby care can be safely transferred to a suitably equipped centre in good time. This improves the outcome for the baby and can avoid stressful situations, such as the mother and baby being cared for in different centres. Another benefit of the tests is better identification of women who will not develop pre-eclampsia, reducing unnecessary hospitalisation. The committee considered that this was particularly relevant during the COVID-19 pandemic to help reduce spread of the virus. Clinical experts also highlighted the benefits of using the tests for shared decision making, with test results helping discussions.
Clinical effectiveness
PLGF-based test results should be used alongside clinical information for decision making
3.2 The committee considered the PARROT and INSPIRE trials, which assessed PLGF-based tests as part of a clinical algorithm that included using the tests alongside clinical judgement to make decisions about care. The PARROT trial was a multicentre, pragmatic, stepped wedge cluster randomised controlled trial of 1,023 women with suspected pre-eclampsia who were between 20 weeks and 36 weeks and 6 days of pregnancy. It was done in 11 maternity units in the UK and used the Triage PLGF test. The INSPIRE trial was a prospective, interventional, parallel-group, randomised clinical trial of 370 women with suspected pre-eclampsia who were between 24 weeks and 37 weeks of pregnancy. It was based in a single UK tertiary referral hospital and used the Elecsys immunoassay sFlt1/PLGF ratio. Clinical experts said that the tests are not a substitute for clinical assessment. Instead PLGF-based testing gives the clinician more evidence to help them make an informed decision. Clinical experts also explained that a low PLGF test result does not always mean a woman has pre-eclampsia and can be associated with other conditions affecting the placenta. They did however highlight that PLGF-based test results can be very useful to help with clinical decision making, particularly for women who had hypertension or proteinuria before becoming pregnant. The committee concluded that PLGF-based test results should be used alongside clinical information for decision making.
PLGF-based testing did not lead to unnecessary early births in UK trials
3.3 In NICE diagnostics guidance 23 (DG23), the committee was concerned that too much emphasis might be placed on PLGF test results indicating pre-eclampsia, which could result in the unnecessary early birth of the baby. Since this guidance was published, the NICE guideline on hypertension in pregnancy has been updated to include recommendations on deciding the timing of birth in women with pre-eclampsia. In the PARROT trial the proportions of births before 37 weeks in the test and control arms of the trial were similar. In the PARROT, PARROT Ireland (a multicentre, pragmatic, stepped wedge cluster randomised controlled trial done in 7 maternity units throughout Ireland) and INSPIRE trials, weeks of pregnancy before birth were also similar in the test and control arms. Clinical experts said that this reflects current practice because the tests are used to help with decisions about hospitalisation and whether to transfer to a specialist unit, not to guide decisions about birth. They also pointed out that about half the centres that participated in the PARROT trial were not specialist centres, which reduced concern about how the tests would be used if they were adopted more widely. The committee concluded that there was evidence that use of the tests did not lead to unnecessary early births.
Maternal outcomes: evidence from trials suggests potential improvements with PLGF-based testing and better decisions about care
3.4 The PARROT trial data suggested that using a PLGF test improved maternal outcomes. In PARROT the number of women with adverse outcomes, defined by the fullPIERS consensus, was lower in the revealed group (4%) than the concealed group (5%), and this difference was statistically significant. Incidence of placental abruption and severe pre-eclampsia was also lower with test use in the INSPIRE trial but not statistically significantly so. The clinical experts explained that the INSPIRE trial was not powered to detect differences in the adverse maternal outcomes that it assessed. In the INSPIRE trial the proportion of women who had confirmed pre-eclampsia within 7 days of testing who were admitted to hospital was greater in the test use arm of the trial (100%) than the control arm (83%). At the second committee meeting, the committee considered the PARROT Ireland randomised controlled trial. It found that integrating PLGF testing into routine clinical investigations for women with suspected preterm pre-eclampsia had no significant effect on maternal morbidity. Clinical experts highlighted differences between the PARROT and PARROT Ireland study cohorts. PARROT Ireland had a higher proportion of women with suspected fetal growth restriction (55%) than PARROT (16%). Also, the incidence of pre-eclampsia in PARROT was higher (35%) than in PARROT Ireland (14%). They also noted that PARROT Ireland had only recruited just over half the proposed participants (2,313 out of a planned 4,000, or 57%) and may have been underpowered to detect significant differences. The committee concluded that there was some evidence that PLGF-based test use could improve management decisions and clinical outcomes for women with suspected preterm pre-eclampsia, although there was considerable uncertainty about this.
Neonatal outcomes: the effect of PLGF-based testing is uncertain but some evidence suggests it may improve decisions about care
3.5 Incidence of perinatal and neonatal mortality and complications in the test and control arms of the PARROT trial were similar. Clinical experts explained that there was a very low number of these clinical events and that the trial was not powered to show differences. The committee considered that it was uncertain whether the differences were down to test use or chance. Clinical experts explained that intraventricular haemorrhage (IVH) and respiratory distress syndrome (RDS) can be devastating for babies and their families. But because they happen rarely, it is difficult to do trials to assess how tests affect them. A clinical expert said that, in a stratified analysis of the PARROT data, more women with a PLGF test result of less than 12 pg/ml who delivered before 35 weeks of pregnancy were given antenatal corticosteroids 7 days before birth in the revealed group (39%) than the concealed group (16%). They explained that this meant the women who had the PLGF test had better clinical care because antenatal corticosteroids reduce the risk of RDS, IVH and death in preterm babies. A clinical expert also pointed out that the number of nights that babies spent in the intensive care or high-dependency unit was only 15.2 nights in the test arm of PARROT, compared with 24.2 nights in the control arm. The committee concluded that, because of the rarity of IVH, RDS and death, the effect of using PLGF-based tests on neonatal outcomes is uncertain. But it agreed there was some evidence that they influence management decisions that could improve care.
The DELFIA tests have established rule in and rule out thresholds, but the BRAHMS test does not
3.6 In DG23, the committee did not recommend the DELFIA Xpress PLGF 1‑2‑3 test or BRAHMS sFlt‑1 Kryptor/BRAHMS PLGF plus Kryptor PE ratio test for routine adoption in the NHS. It said that further research should be done by the companies to show their clinical effectiveness, including diagnostic accuracy and analytical validity. No new data was found for how these 2 tests affect management decisions or clinical outcomes (such as maternal or neonatal outcomes). However, there was new evidence on test accuracy. Since DG23, rule in and rule out thresholds have been established for the DELFIA Xpress PLGF 1-2-3 test based on performance of the test compared with the Triage PLGF and Elecsys immunoassay sFlt-1/PLGF ratio tests (McCarthy et al. 2018 and Giblin et al. 2020). A prospective study using these preset thresholds was also considered at the second committee meeting (Bremner et al. 2022). A quality assessment of this study indicated a low risk of bias and no applicability concerns. This study also provided accuracy estimates for the DELFIA Xpress sFlt-1/PLGF 1-2-3 ratio assay, using specified thresholds. The external assessment group (EAG) identified 2 studies that compared the BRAHMS sFlt-1 Kryptor/BRAHMS PLGF plus Kryptor PE ratio with the Elecsys test. One indicated highly correlated results (Salahuddin et al. 2016) but the authors of the other study (Cheng et al. 2019) indicated that results from the different manufacturers' immunoassays were significantly different, and that sFlt-1/PLGF rule in and rule out criteria are manufacturer specific, not interchangeable, and require separate clinical validation. The committee noted that the manufacturer only specified a rule in threshold for the BRAHMS test. It heard from the company that it intends to specify a rule out threshold in the future. The committee concluded that there was now some evidence on the accuracy of the DELFIA tests, which could address the request for further data in DG23.
Repeat PLGF-based testing evidence is still limited
3.7 DG23 made a research recommendation on using repeat PLGF‑based testing. Not much more evidence was found for repeat PLGF testing for this assessment, but the clinical experts pointed out that ongoing work, for example the PARROT 2 trial, will provide further data in the future. The committee concluded that the research recommendation made in DG23 about repeat testing should be retained in this guidance (see section 4.2).
Cost effectiveness
The DSU model is suitable for decision making
3.8 At the first committee meeting, the committee was concerned about the EAG's model and approach to modelling. The standard assessment costs and quality-adjusted life years (QALYs) in the EAG's model were different for the Elecsys immunoassay sFlt-1/PLGF ratio and the Triage PLGF Test. The EAG explained this was because it used data from the control arm of the INSPIRE trial for the Elecsys test and from the PARROT trial for the Triage PLGF Test. But clinical experts said that there were important differences between the 2 trial populations, for example pre-eclampsia incidence. The committee noted that using different populations to assess standard assessment for different tests made interpreting results more difficult, and could lead to a biased assessment. The level of pre-eclampsia in the test use and non-test use arms of the individual models was also different, particularly for the Elecsys model. This was because the EAG used unadjusted data from the trials, and the pre-eclampsia incidence was higher in the test use arm. A clinical expert said that this was caused by chance allocation to trial arms. There was also uncertainty about whether the populations modelled accurately represented women with suspected preterm pre-eclampsia who would have the PLGF-based tests in the NHS. The committee would have preferred the same model for standard assessment to be used for all tests, and for the model to be based on a population that accurately represents women with suspected preterm pre-eclampsia in the NHS. The committee concluded that more work on the model was needed to address these concerns before any recommendations could be made. As a result of these concerns, NICE commissioned the DSU to carry out further modelling work. For the second committee meeting, the DSU provided an updated model and analyses. The committee said that the DSU's model addressed its previous concerns and concluded that it was suitable for decision making.
It is appropriate to consider cost effectiveness estimates of the tests when used to help rule out and rule in pre-eclampsia
3.9 The DSU provided cost effectiveness estimates when the PLGF-based tests were used to rule out pre-eclampsia only, and when the tests were used to rule in and rule out pre-eclampsia. The committee recalled that data from recent studies provided reassurance that using positive results from the tests to inform care did not lead to earlier births (see section 3.3). Clinical experts said that training and education on these tests focuses on using them to identify women who have a higher risk of developing pre-eclampsia, rather than as a trigger for offering an early birth. The committee concluded that, provided the tests are used alongside clinical judgement (see section 3.2) and are not used to decide on timing of birth (see section 3.14), it is appropriate to use the tests to help diagnose pre-eclampsia. Therefore, it was appropriate to consider cost effectiveness results from the DSU's model in which the tests were used to help rule in and rule out pre-eclampsia.
The Elecsys and Triage tests used to rule out and rule in pre-eclampsia are cost effective
3.10 When used for rule in and rule out in the DSU's base case analysis, testing using the Elecsys immunoassay sFlt-1/PLGF ratio or Triage PLGF test typically dominated standard assessment (that is, they led to lower costs and provided more QALYs). Tests were less cost effective when neonatal outcomes were removed from the model, but incremental cost effectiveness ratios (ICERs) only increased to above £20,000 per QALY gained when decisions about care based on test results were based on the PreOS trial (a multicentre, prospective, open-label, non-interventional study in 150 women with suspected pre-eclampsia) and standard assessment was modelled based on the INSPIRE trial. The committee concluded that the Elecsys and Triage tests were cost effective, compared with standard assessment, when used to rule out and rule in pre-eclampsia.
The DELFIA Xpress PLGF 1-2-3 used to rule out and rule in pre-eclampsia is cost effective
3.11 Costs effectiveness estimates for the DELFIA Xpress PLGF 1-2-3 used to rule in and rule out pre-eclampsia, compared with standard assessment, from the DSU's model were similar to those for the Elecsys immunoassay sFlt-1/PLGF ratio and Triage PLGF test (see section 3.10). The committee recalled that the thresholds for this test were set based on giving the same accuracy as the Triage and Elecsys tests (see section 3.6). The DSU used data from the COMPARE study (a secondary analysis of samples from 3 prospective cohort studies, including 396 women with suspected pre-eclampsia or babies suspected to be small for gestational age, before 35 weeks and between 35 and 36 weeks of pregnancy) for the DELFIA Xpress PLGF 1-2-3 in its model. It noted that this study had no prespecified threshold, which was a concern. Comments received on the DSU report included reference to a recently published prospective study (Bremner et al. 2022) that provided further diagnostic accuracy evidence for the DELFIA Xpress PLGF 1-2-3 test using prespecified cut-offs. The committee was satisfied that there was enough data to show how well the test worked. It concluded that the DELFIA Xpress PLGF 1-2-3 was cost effective, compared with standard assessment, when used to rule out and rule in pre-eclampsia.
The DELFIA Xpress sFlt-1/PLGF 1-2-3 ratio is cost effective when used to rule out and rule in pre-eclampsia
3.12 The Bremner et al. (2022) study (see section 3.11) also provided accuracy estimates for the DELFIA Xpress sFlt-1/PLGF ratio. Because this study had accuracy estimates from the DELFIA Xpress PLGF 1-2-3 alone from the same population, the DSU was able to include this test in its model. The results for this test, compared with standard assessment, were similar to the DELFIA Xpress PLGF 1-2-3 alone. When the DELFIA sFlt-1/PLGF ratio and DELFIA Xpress PLGF 1-2-3 alone were compared with each other, rather than with standard assessment, the DELFIA sFlt-1/PLGF ratio was in general dominated (that is, it had higher costs and produced fewer QALYs). But the committee noted that differences in costs and QALYs were small, and that there was uncertainty about their relative cost effectiveness. The committee also noted that the cost of doing the DELFIA sFlt-1/PLGF ratio is higher than the DELFIA Xpress PLGF 1-2-3 alone, and questioned whether commissioners would want to use the more expensive test without evidence of benefit. Clinical experts said that there may be some additional benefit to including a measurement of sFlt‑1 because it may improve test performance. The committee concluded that the DELFIA Xpress sFlt-1/PLGF 1-2-3 ratio was cost effective, compared with standard assessment, when used to rule out and rule in pre-eclampsia. It said that commissioners could make a decision to add the DELFIA sFlt-1 assay to the DELFIA PLGF 1-2-3 assay, based on locally available costs.
There is not enough data to recommend the BRAHMS Kryptor ratio test and it's not clear how the test is intended to be used
3.13 For the BRAHMS sFlt‑1 Kryptor/BRAHMS PLGF plus Kryptor PE ratio test, the DSU's model used accuracy estimates from the Simon et al. (2020) study. The committee noted that this study used a threshold of 38 for rule out that was not specified by the manufacturer (see section 3.6) This was also a case control study that was not done in the UK and included participants from a high-risk population that were already known to have pre-eclampsia or fetal growth restriction, rather than a population with suspected preterm pre-eclampsia. The committee noted that for this reason the EAG had excluded this study from its original report. The committee said that it had concerns about the size and case control design of the Simon et al. study. Because the company did not have a rule out threshold for the test, clinical experts also said that there would be uncertainty about how the test should be interpreted. The committee concluded that there was too much uncertainty about the diagnostic performance of the BRAHMS ratio test to recommend routine adoption, and further research was needed (see section 4.1).
PLGF-based tests should not be used to make decisions about timing of birth in women with pre-eclampsia
3.14 The committed recalled that data from trials had reassured it that using the tests to help rule in pre-eclampsia had not led to unnecessary early births. Clinical experts emphasised that if the tests were used more widely, it was important that they were not used to make decisions about timing of birth. The committee concluded that it was important to highlight this in the recommendations.
Research considerations
Research is needed on test cut-offs for women pregnant with more than 1 baby
3.15 The INSPIRE and PARROT studies only included women pregnant with 1 baby. Clinical experts said that PLGF or sFlt-1 levels may differ in pregnancies with more than 1 baby because of increased placental mass. Therefore specialists using the tests in this group would interpret the results with caution and potentially not use the specified cut offs. They said that research is needed to find out if different cut-offs are needed.
There is no international standard reference material for PLGF testing
3.16 The committee noted that there is currently no international standard or reference method procedure for PLGF or sFlt-1 testing. This is important for the external quality assurance of laboratories offering this testing, and the committee encouraged the development of such standards.